HPLC – High Performance Liquid Chromatography – Calibration Procedure

HpLC calibration

 

HPLC Calibration

 


SCOPE :

This Procedure is extended to the High Performance Liquid Chromatograph instruments.

 

PROCEDURE :

 

Cleaning of flow cell :

To achieve satisfactory noise and drift characteristics of the detector a flow cell of the detector is cleaned by passing solvent through the flow cell in the following order :

Column should be replaced by union during this treatment.

SOLVENTS USED

VOLUME

Deionised water     

20 ml

10 % NH4OH/ Water Solution

20 ml

Deionised water

20 ml

Nitric Acid ( 20 % v/v)        

20 ml

Deionised water

20 ml

Isopropyl Alcohol

20 ml

Deionised water

20 ml

Pump Calibration/ Flow Rate Accuracy:

 

Connect Inert sill ODS, 250 mm X 4.6 mm HPLC Column and calibrate the pump flow at  1 ml per minute flow rate.  

To check the flow rate accuracy at 5 ml per minute, disconnect the column and connect the guard column 10GLC4- C8, 10mmX 4 mm and calibrate the pump.

Pump Flow Composition of Mobile
Phase
Time Limit Collection Volume
1.0 ml/ min Water: Methanol (1:1) 25 min 24 to 26 ml
5.0 ml/ min Water: Methanol (1:1) 10 min 48 to 52 ml

Detector Calibration/ Wavelength Accuracy :

 

Prepare a 0.02 mg/ ml (0.002 % i.e. 20 ppm) solution of caffeine in Mobile phase. 

 

Run a chromatogram of above caffeine solution, using following Chromatographic Conditions.  

Column

Inert sill ODS, 250 mm X 4.6 mm
(5 micron)

Mobile Phase          

Water: Methanol (70:30)

Flow Rate

1 ml/ min

Injection Volume    

20 micro lit.

Chromatogram should be recorded at each of the following wavelength

Sr. No.

Wave Length ( nm )

1

238

2

240

3

242

4

244

5

246

6

268

7

270

8

272

9

274

10

276

 

Plot a graph of absorbance versus wavelength and find out the absorbance maxima and minima.
Limit for Maxima Limit for Minima
272 nm ±  2 nm 242 nm ±  2 nm

 

HPLC - High Performance Liquid Chromatography - Calibration

 

Injection Reproducibility:

Calibration Solution:

Stock Solution: Prepare 0.75 mg/ ml ( 0.075 % i.e. 750 ppm ) Caffeine Solution in mobile phase.
Prepare a calibration solution by diluting 15 ml stock solution to 100 ml with mobile phase.
 
Carry out the test on calibration solution by using following Chromatographic Conditions.  
 

Column

Inert sill ODS, 250 mm X 4.6 mm
(5 micron)

Mobile Phase          

Water: Methanol (70:30)

Flow Rate

1 ml/ min

Detector wavelength

272 nm

Injection Volume    

20 micro lit.

Record the 6 successive injection chromatogram and determine the Relative Standard Deviation (RSD). 
 
Limit : RSD should be less than 2 %

PRECAUTIONS : 

  • Use only HPLC grade solvents
  • Use filtered and degassed mobile phase and water.

Calibration Frequency- Every Six Month